BrainGate analysis tracks Utah array performance across 20 years

A post hoc analysis of BrainGate and BrainGate2 data assessed 20 intracortical arrays in 14 participants across 2,289 recording sessions. Most arrays maintained useful movement-decoding signals during participation, although recording yield varied and declined modestly on average.

Researchers conducted a post hoc analysis of participants enrolled in the BrainGate and BrainGate2 pilot trials from 2004 to 2019. Fourteen people received one or two 96-channel Utah arrays. The analysis covered 2,289 recording sessions and 20 arrays; decoding performance was assessed in 729 sessions, with high-bandwidth data available for a subset of participants. Outcomes included electrode recording yield, cursor-movement decoding and signal stability over time.

Across arrays, an average of 35.6% of electrodes recorded neural spiking waveforms, with a 7-percentage-point decline over the recording period, which extended to as long as 7.3 years. Eleven of 14 arrays had a median decoding signal-to-noise ratio above 1. Three achieved cursor-control accuracy above 70% of the accuracy recorded for able-bodied mouse control. Decoding signal-to-noise ratio increased approximately logarithmically with electrode number.

The findings provide a multi-participant benchmark for intracortical BCI performance and suggest that useful movement-decoding signals can persist for years. However, performance varied between arrays, peak performance was not consistently achieved, and decoding analyses were limited by available high-bandwidth data and task sessions. The analysis included the first 14 trial participants and did not include those enrolled from 2021 to 2025; it does not establish clinical effectiveness for broader populations.

In practice

For researchers and clinicians evaluating intracortical BCIs for people with paralysis, the analysis indicates that useful cursor-movement decoding signals can persist across years of participation, but performance is heterogeneous: 11 of 14 arrays had median decoding signal-to-noise ratios above 1, and only three exceeded 70% of able-bodied mouse-control accuracy. These retrospective trial data do not establish effectiveness in broader populations or support a change in clinical practice. Decoding analyses were limited to available high-bandwidth recordings and sessions, and participants enrolled after 2019 were not included.

AI-generated clinical interpretation · Not editor-reviewed

AI summary · Not yet editor-reviewed

Is this summary clinically accurate?

Help fellow clinicians: your rating sends inaccurate summaries straight to our editors.

Sign in to rate this summary →

Source

Nature Medicine: Performance of intracortical microelectrode arrays in people with implanted brain–computer interfaces over a 20-year period ↗

This is an automated AI-condensed summary that has not yet been reviewed by an editor. Always consult the full item at the original source.